RPFT Exam Question No 39

Question No 39:

An isothermal lung analog has the following specifications:

5-L glass bottle
2000 g of steel wool
Density of steel wool 5 g/cm3

Volume of connectors and bulb 100 mi-Repeated FRC measurements in the plethysmograph
result in a volume of 5.5 L. Which of the following should a pulmonary function technologist
do?

A.
Remove 500 g ofsteel wool from the lung analog
B.
Take the plethysmograph out of service
C.
Recalculate the BTPS correction
D.
Begin a biologic control program

Answer: D